This is how we set up our Lumina button response box with Presentation using the parallel port.
Hardware:
- get an RJ45 to D25 connector (e.g. from Minitran)
 - connect pin 5 (green cable; trigger pulse) with pin no. 10 on parallel port
 - connect pin 8 (white/grey cable; ground) with pin no. 24 on parallel port
 
Use hyperterminal to configure the Lumina response box:
- start hyperterminal
 - click cancel
 - File > Properties
 - connect using COM1
 - set:
    
- baud rate = 9600
 - data bits = 8
 - stop bits = 1
 - flow control = none
 
 
Settings Lumina response box:
- ASCII/MEDX
 - 9600 baud
 
Configure Presentation:
- In Presentation, select the Settings tab and then select Response on the left.
 - Click the Parallel Port Device button:
 

 - On the Parallel Input Device dialogue box, enter the following settings.
 

 - Click Close.
 - Click the Serial Input Device button:
 - On the Serial Input Device Settings dialogue box, enter the following settings.
 

 - Click OK.
 - Select Port on the left, and enter the following settings:
 

 